Temperature Sensors With Fittings And Fixtures from Ellab A/S Denmark For Precise Retort Product Validation (Heat Penetration Study)

Retort Validation (TD and HP Studies) should Not to be done for the Sake of Doing it. It should be done perfectly and precisely using internationally approved and accepted validation equipment in compliance with FDA 21 CFR Part 11 (to avoid data tampering / manipulation), with highly reliable and accurate sensors with proper heat penetration fittings and fixtures.
This is because the Heat Penetration Study in conjunction with the corresponding TD study is the Heart of the Retort Processing of any products. The retort validation studies decide the Safety and Quality of the Retort Food products, because regular commercial production fully depends on the control parameters derived from the validation studies. So processors SHOULD NOT adjust, bypass or compromise the critical parameters involved in the retort validation activities.
A systematic or experimental error of 1 deg C in a temperature measurement system at the sterilization reference temperature of 121.1 deg C, would lead to a corresponding error of 26% in the calculated Fo Sterility Value, resulting either under-sterilization or overcooking of the products.
This is because the Heat Penetration Study in conjunction with the corresponding TD study is the Heart of the Retort Processing of any products. The retort validation studies decide the Safety and Quality of the Retort Food products, because regular commercial production fully depends on the control parameters derived from the validation studies. So processors SHOULD NOT adjust, bypass or compromise the critical parameters involved in the retort validation activities.
A systematic or experimental error of 1 deg C in a temperature measurement system at the sterilization reference temperature of 121.1 deg C, would lead to a corresponding error of 26% in the calculated Fo Sterility Value, resulting either under-sterilization or overcooking of the products.
Critial Points Involved in Heat Penetration Study

Correct positioning of the sensor measuring tip with the core of the sample product, at the cold spot of the container.

Elimination of Steam and Water ingression during measurement.

Minimize the Sensor Assembly inside the containers to minimize the risk of heat conduction.

Simulate Actual Production Conditions inside the containers during measurement.

Ellab offers perfect Retort Validation Equipment with Software, in compliance with FDA 21 CFR Part 11, involving sensors with an unmatched accuracy of +/-0.05 deg C | C/ 0.1 deg C with proper fittings and fixtures compatible with different types of packing containers to perform accurate, precise and reliable Heat Penetration Studies.

You can see here how Ellab sensors and HP fittings can be used to perform proper Heat Penetration Studies.
Read the complete heat penetration guide.
You can see here how Ellab sensors and HP fittings can be used to perform proper Heat Penetration Studies.














Please contact us for any further clarifications or technical assistance.
How Can We Help You?
Connect with our dedicated team to explore tailored solutions for your inquiries. Whether you're seeking expert advice or have specific questions, we're here to assist you. Our technical experts are committed to delivering prompt and precise responses, ensuring you receive the perfect solution for your needs.